Triangle Theory

SMB chromatography design using the Triangle Theory for separation region analysis

Example Presets
Column
Feed Concentration
Isotherm
Liquid Flow
No separation

m₁ = 1,444.01

m₂ = 941.7452 (need 0.701980–0.715842)

m₃ = 753.3962 (need 0.701980–0.715842)

m₄ = 1,004.53 (need < 0.701980)

HA = 0.715842, HB = 0.701980

Zone Parameters
Zone 1Q = 23.0000 L/minu = 1.8050 cm/s
Zone 2Q = 15.0000 L/minu = 1.1772 cm/s
Zone 3Q = 12.0000 L/minu = 0.941745 cm/s
Zone 4Q = 16.0000 L/minu = 1.2557 cm/s
Vcol = 0.265465 L | εt = 0.700000 | Qsolid = 0.015928 L/min
Triangle Diagram (m₂ vs m₃)
HBHAHBm₃m₂0.396683216.9587433.5208650.0829866.64491,083.210.396683216.9587433.5208650.0829866.64491,083.21(753.3962, 941.7452)